Signal cables and a second peak in the decay

Understanding-oriented. A bench test on 24 and 25 November 2026: the TCSPC histogram of a 2.5 ns calibration slab showed a second, smaller peak after the main one. Setup: FLIM LABS card, 488 nm pulsed laser, galvo scanner on the 7 × 4 plate, Thorlabs long-pass emission filter, PLF reconstruction. The cause was the coax cable from the detector to the card.

In short​

  • The second peak is an echo of each photon pulse on the cable from the detector to the card. The card counts the echo as a second photon, one cable round trip later.
  • Its delay followed the cable: 12.2 ns with the long cable, 6.4 ns with the short one. A new cable removed it.
  • Two cables joined by a barrel made the echo almost as large as the photon pulse: every photon was counted twice.
  • The laser, the optics and the objective were not involved: changing the repetition rate, the objective and the excitation filter did not move the second peak.
  • A cable change also moves the whole decay in time. Calibrate again after changing any cable.

What the second peak looked like​

On the 2.5 ns slab the main peak decays as expected. A second peak with the same shape sits a fixed time later and is 8-10 % as high:

TCSPC histograms of the 2.5 ns slab with different cables and settings. a-c: long cable at 40 and 20 MHz, in daylight and covered: second peak 12.2 ns after the main one. d-f: short cable: 6.4 ns, also with another sync cable and with a 490 nm short-pass after the laser. g: two cables joined by a barrel: the second peak is almost as high as the first. h: new cable: no second peak

The delays and heights were read from the screenshots (about ±0.3 ns):

testlasersecond peak after the main oneheight of the second peak
long cable, daylight40 MHz12.0 ns9 %
long cable, daylight20 MHz12.4 ns9 %
long cable, covered20 MHz12.2 ns9 %
long cable80 MHznot visible: 12.2 ns is one period (12.5 ns), so it lands on the main peak-
short cable40 MHz6.4 ns10 %
short cable, other sync cable40 MHz6.3 ns9 %
short cable, other objective40 MHz6.4 ns9 %
short cable, SP490 short-pass after the laser40 MHz6.3 ns8 %
two cables joined by a barrel40 MHz6.4 ns89 %
new cable40 MHznone-

How the tests narrowed it down​

suspecttestresultverdict
a second laser pulse half a period later20 and 80 MHz instead of 40 MHzthe delay stayed at about 12.2 ns, not half the period (25 ns at 20 MHz)not the laser
a reflection in the opticsanother objective; geometryno change; a delay of 12 ns needs 3.7 m of free-space path, the beam path on the plate is well under 1 mnot the optics
daylightcover on (20 MHz)the flat floor under the decay dropped from 12 % to 4 % of the peak, the second peak stayed at 9 %daylight makes the floor, not the peak
broad laser emission490 nm short-pass after the laserthe second peak stayed (8 %)not the laser spectrum
the sync cable from the laser to the cardanother sync cablethe whole decay moved 1.5 ns earlier, the delay between the peaks stayednot the sync cable
the cable from the detector to the carda shorter cablethe delay halved, from 12.2 to 6.4 nsthis cable
two cables joined by a barrelsecond peak 89 % as high, count rate about twice as high
a new cableno second peak

Why a cable makes a second peak​

Each photon produces a short electrical pulse at the detector. The pulse runs down the coax to the card. Where the cable meets something that does not have the same impedance (the card's input, a barrel connector, a cable of a different type), part of the pulse is reflected and runs back. At the detector it is reflected again and arrives at the card one round trip later.

In a cable with solid polyethylene insulation (RG-58, RG-174, RG-316) a pulse travels about 20 cm per nanosecond, so a round trip takes about 10 ns per metre of cable:

delay of the second peakcable length for that round trip
12.2 nsabout 1.2 m
6.4 nsabout 0.64 m

The card does not see a difference between a photon pulse and its echo. Whenever the echo is still above the card's threshold, it is counted as one more photon, a round trip later. With the old cables that happened for 8-10 % of the photons. A barrel between two cables reflects much more strongly, so nearly every pulse came back above the threshold: two equal peaks, and the count rate doubled (2.77 M counts/s, against 1.43 M counts/s with the new cable shortly after).

The new cable most likely has the impedance the detector and the card expect (50 Ω), so almost nothing is reflected.

✏️ TODO - Benedict

Add the type and length of the old long, old short and new detector cable, and the detector model. The delays above predict about 1.2 m and 0.64 m for the two old cables.

What the echo does to the lifetimes​

  • Intensity. Each echo is counted as a photon, so the intensity is 8-10 % too high, and twice as high with the joined cables.
  • Phasors. The echo adds a delayed copy of the decay, which pulls the phasor of every pixel inside the universal circle. A calibration with the 2.5 ns slab corrects this, as long as the echo stays the same between the calibration and the measurement. It does not if a cable is changed, re-plugged or bent differently.
  • Fits. A tail fit that does not know about the echo reads the second peak as part of the decay.

A cable change also moves the whole decay: with the other sync cable the main peak arrived 1.5 ns earlier. At 40 MHz that is a phase change of 22°, a large lifetime error if the old calibration is still used. Calibrate again after changing or re-plugging any cable, the detector cable and the laser sync cable alike.

Why the phasors were scattered​

Phasor plots of the 2.5 ns slab with the short cable, 4 frames: a cloud from near the 2.5 ns point on the universal circle towards the origin

A 2.5 ns lifetime at 40 MHz belongs at g = 0.72, s = 0.45 on the universal circle. The clouds above run from there towards the origin, with a lot of spread. Two things cause that, not the cable:

  • Few photons per pixel. At about 600 000-870 000 counts/s, a 12.5 µs dwell and 4 frames, a pixel collects only 30-45 photons. That gives a phasor spread of roughly ±0.1 in g and s. About 1000 photons per pixel bring it down to about ±0.02: more frames, a longer dwell, or binning pixels (Bins in FLIM Imager).
  • Background in dim pixels. Light that does not come from the laser pulse (daylight, detector dark counts) has no lifetime and pulls a pixel's phasor towards the origin. The dimmer the pixel, the stronger the pull. The edges of the field are dimmer than the centre, so the cloud stretches towards the origin. Measure covered, and leave out dim pixels with an intensity threshold.

Choosing and checking the cables​

  • Use 50 Ω coax with the connectors of the detector and the card (SMA) at both ends. Do not use 75 Ω video or TV cables, or cables of unknown type.
  • Use one continuous cable per signal: no barrels, adapters or T-pieces in the line.
  • Keep the cables short, and keep them the same from day to day. Label the detector cable and the sync cable.
  • After any change, check a mono-exponential sample (the calibration slab) in the TCSPC view: one peak, no bump after it, and the count rate as before.
  • Calibrate again after any cable change.
  • The flat floor under the decay is light that is not timed to the laser: measure with the cover on and the room light off. In these tests daylight raised it from 4 % to 12 % of the peak at 20 MHz, and to 31 % at 40 MHz.
  • A short-pass (here SP490) after the laser blocks the laser's weaker emission at longer wavelengths, which the emission filter would let through. It did not affect the second peak.