Skip to main content

Controlling the Smearscope

The Smearscope has three devices that you talk to over USB:

DeviceConnectionWhat it does
QuadDriverUSB-C (virtual serial port) or CANmoves the XY stage
LED matrix (ESP32-S3)USB-Csets the illumination pattern
Hikrobot cameraUSB 3takes the images

Focus is set by hand with the knob under the stage.

Moving the stage​

The QuadDriver appears as a USB serial port (USB CDC). It takes plain-text commands, one per line, and answers OK, a value, or ERR …. Any serial terminal works, for example the Arduino serial monitor, PlatformIO's device monitor, or Python with pyserial. Type HELP to get the command list.

Distances are in microsteps. At the default 32 microsteps per electrical cycle, one microstep is about 130 µm and one cycle (32 microsteps) about 4.16 mm.

Motion commands​

CommandEffect
MOVE <x> <y> [speed]move relative, in microsteps, at speed microsteps/s (default 500; 100 is recommended)
V <vx> <vy>run continuously at the given velocities (microsteps/s)
STOPstop (same as V 0 0)
SNAKE <nx> <ny> <sx> <sy> <speed> <pause_ms> [hold%]serpentine scan: nx × ny positions, steps sx, sy, pause at each position
POSstep counters of both axes, <x> <y>
BUSY1 while the stage is moving, else 0
PINGanswers OK

MOVE answers OK at once and runs in the background. Poll BUSY until it returns 0 before you take an image.

Settings​

CommandEffectRecommended
POWER [x y]peak drive per axis, 0–3276718000 24000 (the Y coils have higher resistance)
ACCEL [n]acceleration limit, microsteps/s²1500 (default)
HOLD [pct]coil current when the stage stands still, 0–100 %0: holding current adds friction and heat
MICROSTEP [n]16, 32 or 64 microsteps per cycle (stored, applies after a restart)32
DITHER [mode amp freq]small modulation of the drive to help break friction1 16 50 or off
PHASE <X|Y> <k> [v]fine correction of the commutation tableleave at zero

POWER, ACCEL, DITHER and PHASE are not stored and reset on power-up.

Example in Python​

import serial, time

stage = serial.Serial("COM5", 115200, timeout=1) # macOS/Linux: /dev/cu.usbmodem… or /dev/ttyACM0

def cmd(line):
stage.write((line + "\n").encode())
return stage.readline().decode().strip()

def move(dx, dy, speed=100):
cmd(f"MOVE {dx} {dy} {speed}")
while cmd("BUSY") == "1":
time.sleep(0.02)

cmd("POWER 18000 24000")
move(32, 0) # one electrical cycle in +X, about 4.2 mm
print(cmd("POS")) # e.g. "32 0"

Tips for reliable motion​

  • Move in whole electrical cycles (multiples of 32 microsteps) when you can. These are the most repeatable steps. Single microsteps are not reliable positioning steps.
  • Approach positions from one direction. Each reversal loses about 0.2–0.4 mm to friction. For scans, run every row in the same direction, or overshoot and come back.
  • Keep the speed at about 100 microsteps/s. Much faster profiles can make the sled lose step.
  • Keep holding current off. Current in the coils while standing still adds friction, so the next move starts worse.
  • Let the sled settle. A move ends with a short slip. Wait about 50 ms after BUSY returns 0 before taking an image.

CAN bus​

The QuadDriver can also join an openUC2 CAN bus (CANopen, 500 kbit/s, default node ID 11). Velocity, acceleration, drive level, holding current and microstep mode are exposed as CANopen objects. The CAN interface is intended for integration into larger openUC2 systems.

Illumination​

The LED matrix is the same ESP32-S3 8 × 8 RGB matrix used in other openUC2 kits. It is set with commands from the computer, for example a pattern of LEDs and a colour. The How it works page lists the useful patterns.

Camera​

The Hikrobot camera works with the vendor's MVS software and with ImSwitch. See the Hikrobot camera documentation for drivers and settings.

Scanning and stitching​

A typical slide scan works like this:

  1. Focus on the sample with the knob.
  2. Choose a grid with at least 30 % overlap between neighbouring fields.
  3. Step through the grid with MOVE commands, wait for BUSY = 0, and take an image at every position.
  4. Stitch the images by registration: the overlap between neighbours gives their true offsets, and a small least-squares fit gives every field's position.

The camera, not the stage, defines where each field lies. In a 10 × 10 benchmark scan, the stage misplaced fields by up to 0.4 mm, but registration recovered every field's position to about 3 µm and no area was missed.

Stitched 10 × 10 scan from a benchmark run

Benchmark scan: 100 fields in about two minutes, stitched by registration. The dark band at the bottom is an opaque marking on the slide. Example scans of smears will follow.

ImSwitch​