Controlling the Smearscope
The Smearscope has three devices that you talk to over USB:
| Device | Connection | What it does |
|---|---|---|
| QuadDriver | USB-C (virtual serial port) or CAN | moves the XY stage |
| LED matrix (ESP32-S3) | USB-C | sets the illumination pattern |
| Hikrobot camera | USB 3 | takes the images |
Focus is set by hand with the knob under the stage.
Moving the stage
The QuadDriver appears as a USB serial port (USB CDC). It takes plain-text commands, one per line, and answers OK, a value, or ERR …. Any serial terminal works, for example the Arduino serial monitor, PlatformIO's device monitor, or Python with pyserial. Type HELP to get the command list.
Distances are in microsteps. At the default 32 microsteps per electrical cycle, one microstep is about 130 µm and one cycle (32 microsteps) about 4.16 mm.
Motion commands
| Command | Effect |
|---|---|
MOVE <x> <y> [speed] | move relative, in microsteps, at speed microsteps/s (default 500; 100 is recommended) |
V <vx> <vy> | run continuously at the given velocities (microsteps/s) |
STOP | stop (same as V 0 0) |
SNAKE <nx> <ny> <sx> <sy> <speed> <pause_ms> [hold%] | serpentine scan: nx × ny positions, steps sx, sy, pause at each position |
POS | step counters of both axes, <x> <y> |
BUSY | 1 while the stage is moving, else 0 |
PING | answers OK |
MOVE answers OK at once and runs in the background. Poll BUSY until it returns 0 before you take an image.
Settings
| Command | Effect | Recommended |
|---|---|---|
POWER [x y] | peak drive per axis, 0–32767 | 18000 24000 (the Y coils have higher resistance) |
ACCEL [n] | acceleration limit, microsteps/s² | 1500 (default) |
HOLD [pct] | coil current when the stage stands still, 0–100 % | 0: holding current adds friction and heat |
MICROSTEP [n] | 16, 32 or 64 microsteps per cycle (stored, applies after a restart) | 32 |
DITHER [mode amp freq] | small modulation of the drive to help break friction | 1 16 50 or off |
PHASE <X|Y> <k> [v] | fine correction of the commutation table | leave at zero |
POWER, ACCEL, DITHER and PHASE are not stored and reset on power-up.
Example in Python
import serial, time
stage = serial.Serial("COM5", 115200, timeout=1) # macOS/Linux: /dev/cu.usbmodem… or /dev/ttyACM0
def cmd(line):
stage.write((line + "\n").encode())
return stage.readline().decode().strip()
def move(dx, dy, speed=100):
cmd(f"MOVE {dx} {dy} {speed}")
while cmd("BUSY") == "1":
time.sleep(0.02)
cmd("POWER 18000 24000")
move(32, 0) # one electrical cycle in +X, about 4.2 mm
print(cmd("POS")) # e.g. "32 0"
Tips for reliable motion
- Move in whole electrical cycles (multiples of 32 microsteps) when you can. These are the most repeatable steps. Single microsteps are not reliable positioning steps.
- Approach positions from one direction. Each reversal loses about 0.2–0.4 mm to friction. For scans, run every row in the same direction, or overshoot and come back.
- Keep the speed at about 100 microsteps/s. Much faster profiles can make the sled lose step.
- Keep holding current off. Current in the coils while standing still adds friction, so the next move starts worse.
- Let the sled settle. A move ends with a short slip. Wait about 50 ms after
BUSYreturns0before taking an image.
CAN bus
The QuadDriver can also join an openUC2 CAN bus (CANopen, 500 kbit/s, default node ID 11). Velocity, acceleration, drive level, holding current and microstep mode are exposed as CANopen objects. The CAN interface is intended for integration into larger openUC2 systems.
Illumination
The LED matrix is the same ESP32-S3 8 × 8 RGB matrix used in other openUC2 kits. It is set with commands from the computer, for example a pattern of LEDs and a colour. The How it works page lists the useful patterns.
Camera
The Hikrobot camera works with the vendor's MVS software and with ImSwitch. See the Hikrobot camera documentation for drivers and settings.
Scanning and stitching
A typical slide scan works like this:
- Focus on the sample with the knob.
- Choose a grid with at least 30 % overlap between neighbouring fields.
- Step through the grid with
MOVEcommands, wait forBUSY = 0, and take an image at every position. - Stitch the images by registration: the overlap between neighbours gives their true offsets, and a small least-squares fit gives every field's position.
The camera, not the stage, defines where each field lies. In a 10 × 10 benchmark scan, the stage misplaced fields by up to 0.4 mm, but registration recovered every field's position to about 3 µm and no area was missed.

Benchmark scan: 100 fields in about two minutes, stitched by registration. The dark band at the bottom is an opaque marking on the slide. Example scans of smears will follow.
ImSwitch