Skip to main content

Technical reference

All values refer to the current prototype. Nominal values follow from the design. Measured values come from benchmark runs on a prototype stage (see measured stage performance).

Optics​

ItemValue
Configurationinverted, transmitted light, infinity-corrected, beam folded by a 45° mirror
Objective10× / 0.25 plan, RMS thread (0.8"-36), parfocal about 45 mm
Tube lensC-mount lens, f ≈ 100 mm (nominal)
Fold mirrorfirst-surface mirror 40 × 30 × 2 mm
Total magnification5× (nominal, for an objective designed for a 200 mm tube lens)
CameraHikrobot MV-CE060-10UM: 3072 × 2048 px, 2.4 µm pixels, mono, USB 3, C-mount
Pixel size at the sample≈ 0.48 µm (nominal)
Field of view≈ 1.47 × 0.98 mm, 1.77 mm diagonal (nominal)
Resolution limit≈ 1.3 µm (Rayleigh, 550 nm, NA 0.25)
Illumination8 × 8 RGB LED matrix (Waveshare ESP32-S3-Matrix), LED plane about 44 mm above the sample

XY stage​

ItemValue
Principleplanar PCB motor: two orthogonal trace sets, two phase groups each, magnetic sled
Stator board148 × 48 × 1.52 mm, 4-layer FR-4, Ø15 mm optical aperture in the centre
Coil resistance≈ 1.4 Ω (X, outer layer), ≈ 3.2 Ω (Y, inner layer)
Electrical cycle≈ 4.16 mm (measured, X)
Microstepping16, 32 (default) or 64 microsteps per electrical cycle
Step size≈ 130 µm per microstep at 32 microsteps (measured, X)
Recommended speed100 microsteps/s (≈ 13 mm/s commanded)
Supply12 V motor supply, about 30 W per axis at full drive while moving; coils are off at rest
Mounting4 × M3 on a 140 × 40 mm pattern

Stator board dimensions

Focus (z) stage​

ItemValue
Typemanual linear stage (HAJ CTM28H), knob at the bottom
Travel≈ 13 mm (about +4.5 / −8.5 mm from the nominal focus)
Objective mountprinted carrier with an RMS thread

Electronics: QuadDriver​

The QuadDriver board drives the four coil groups of the stage. Its firmware is open source: i4-controller-firmware by Jeff McBride, extended by openUC2.

ItemValue
Size82 × 63 mm, 4 × M2.5 holes on a 76 × 57 mm pattern
MicrocontrollerSTM32G474RE (Arm Cortex-M4F, 128 MHz)
Drivers4 full H-bridges (16 N-MOSFETs, 8 gate drivers), 22 µH series inductor per channel
PWM100 kHz, sine commutation with per-microstep correction
InterfacesUSB-C (USB serial), CAN (CANopen, 500 kbit/s, JST PH 4), SWD (Tag-Connect)
Connectorsmotor supply (screw terminal), 5 V (screw terminal), coil outputs CH1–CH4 (JST PH 8)
FirmwareRust, async RTOS (lilos)

QuadDriver dimensions and connectors

Mechanics and dimensions​

ItemValue
Overall size≈ 200 × 70 × 160 mm (L × W × H, including camera, LED arm and electronics cover)
Sample plane≈ 104 mm above the table
Housingtwo mirrored 3D-printed halves, 50 mm thick together, joined with M3 screws into heat-set inserts
FastenersISO 4762 (DIN 912) M3, M2.5 and M2 socket head screws, M3 × 4 threaded bushings
Printed partsfront half, back half, objective carrier, electronics cover (STL files on the gallery page)

Measured stage performance​

The figures below come from benchmark runs on a prototype stage with the sled sliding on the bare board. Position was measured with a microscope camera by image registration. They describe one board, one sled and one surface, and the friction of the surface changed within a day. Treat them as indicative.

QuantityX axisY axis
Distance per electrical cycle4.15–4.16 mm3.5–3.9 mm
Spread of one full-cycle step± 2–32 µm± 60–840 µm
Return to the start after ± 8 mm with two reversalswithin 4–17 µmwithin 16–74 µm
Loss when the direction reverses220–400 µm260–650 µm
Lowest drive level that moves the sled12 % of full scale37 % of full scale
Time from command to first motion50–100 ms85–110 ms
Peak speed of the sled≈ 23 mm/s≈ 22–26 mm/s

The X axis, on the outer copper layer, is the more reliable one. The Y axis needs more drive and varies more. Moving in whole electrical cycles gives the most repeatable steps. Single microsteps move only 45–70 % of their nominal distance.

Step ladders: position after each step, both axes

Grid scan​

A 10 × 10 scan with 1.04 mm steps and 1.7 mm fields (39 % nominal overlap) took about two minutes.

QuantityValue
Gaps between neighbouring fieldsnone; smallest overlap 31 %
Field position after stitching≈ 3 µm (registration residual)
Placement error of the stage129 µm RMS, largest 435 µm
Main error source≈ 230 µm lost at each row reversal
Angle between the X and Y axes90° ± 0.03°

Measured minus commanded field positions

The stage misplaces fields by up to a few hundred micrometres, but the overlap between fields is always kept. Image registration recovers each field's true position. For surveying a slide, coverage matters more than positioning accuracy, and the camera supplies the accuracy.