FRAME hardware, units ID10–ID20

This section describes the hardware of the FRAME units with serial numbers 10 to 20. It is the technical reference for anyone who builds on, services or evaluates one of these units.

All dimensions come from the CAD of unit 10 and from the FR120 part revision. Measured performance comes from two internal test series in 2026. Where a number is from CAD or a data sheet and not from a measurement, the page says so.

FRAME ID10–ID20, main assemblies

Main assemblies, rendered from the CAD of unit 10. The lamp arm stands at the back; the camera points to the front.

PageWhat it covers
Optical pathOpenings on the axis, tube lens, field of view, sampling, illumination
MechanicsFrame, XY stage, focus drive, objective slider, optics drawer
Electronics and safety functionsBoards, bus, power, emergency stop, laser interlock
Stage repeatability and driftMeasured: open loop against encoder, backlash, drift over 12 to 24 h

What is the same on every unit in this range​

  • Inverted layout: the sample lies on a motorised XY stage, the objectives sit below it on a two-position slider.
  • The sample moves for focus, the objective does not. The optics below the objective stay in a fixed position relative to the objective's back focal plane.
  • A removable optics drawer with 3 x 3 cube positions in one or two layers, below the stage.
  • A slot of 33 mm height between drawer and stage for flat slide-in modules such as the focus lock.
  • One 12 V supply, one CAN bus, one driver board per motor, a Raspberry Pi 5 running ImSwitch.

What differs inside the range​

Two part revisions exist in the CAD. Check your unit before you rely on an opening diameter.

PropertyFR110FR120
Bore in the objective plate14 mm23 mm
Bore in the stage base plate15 mm25 mm
Bore in the intermediate plate20 mm32 mm
Cube layers under the stagetwoone (stage 55 mm lower) or two
Position feedbacknonelinear encoders on X and Y, rotary encoder on Z, none on the objective slider
Crash detection under the objectivenoneoptional thin-film force sensor

Unit 10 was built as FR110 and is the unit on which the encoder tests ran. The parts lists name FR110 for units 7 to 12 and FR120 for the following order; ask openUC2 which revision your serial number carries.

The stack​

The stack on the optical axis

Heights are measured from the top of the base plate. The focus stages stand at the front and back of the Y plate, 178 mm apart; the figure shows them projected onto the axis.

Key data​

Overall dimensions

PropertyValueBasis
Width310 mmCAD
Depth395 mm with the lamp arm; 516 mm with the camera at the frontCAD
Height above the table272 mm to the sample plate, 406 mm to the top of the lamp headCAD
MassNot confirmed. Documents state under 16 kg, 20 kg and 25 kgweigh your unit
XY travel127 x 86 mmmanual
Focus travel8 mm used; 12.7 mm actuator strokemanual, data sheet
XY driveNEMA 11 motor with 1 mm ball screw, 5 µm per full step, 0.3125 µm per microstep (1/16)data sheet
Focus drivetwo NEMA 8 captive actuators, 0.5 mm lead, wedge stage 1:1, 2.5 µm per full stepdata sheet
Objectivestwo, RMS or M25, each on its own parfocal supportCAD
Tube lensf = 180 mm, 23 mm diameterCAD
CameraHikrobot MV-CS060-10UC/UM-PRO, Sony IMX178, 3072 x 2048 pixels of 2.4 µm, 8.9 mm diagonaldata sheet
Fully lit image circleabout 10 mm for objective pupils up to 9 mmray model, see Optical path
Supply12 V DC, 6.67 A (Mean Well GSM90A12-P1M)assembly
ComputerRaspberry Pi 5 with HAT+assembly
BusCANopen, 500 kbit/s, 4-pin JST XH with 12 VElectronics

Design decisions​

Why the sample moves for focus. Structured illumination, confocal and other scanned paths are aligned to the back focal plane of the objective. If the objective moved for focus, that plane would move against the pre-aligned optics, and the illumination would pick up defocus. With a fixed objective, a module can be aligned once on the bench and stays aligned through a focus stack.

The price is a heavier focus axis: the whole sample plate is lifted. The focus section of Mechanics describes the consequences and the planned improvements, including a piezo fine axis.

Why the optics sit in a drawer. The drawer separates from the stand inside the collimated beam. A sideways error of the drawer does not move the image; only a tilt does. The drawer rests on three ball points and is pressed down inside their triangle.

Known limits of this hardware range​

LimitConsequenceStatus
The tube lens is the smallest opening on a 169 to 224 mm long pathSensors larger than 1/1.8 inch are not fully lita larger lens ahead of the fold mirror is planned
FR110 bores of 14 to 20 mmA 4x/0.20 objective works at NA 0.14fixed in FR120
No position feedback as shippedLost steps are not seen; backlash of 1 to 15 µmencoders validated on unit 10
Two independent focus drivesThe sample plate can tilt if the drives lose step against each otherrotary encoder planned
Printed PLA parts in the lamp arm, drawer guides and sensor flagsUse above 38 °C is not supportedmaterial change planned
Transmitted light is diffuse, not KöhlerIllumination NA 0.29; no field or aperture stopby design in this range