Skip to main content

Mechanics (ID10–ID20)

The stand is a stack of machined aluminium plates on four corner columns. The stage plates carry X, Y and focus; the optics drawer slides in below. All values are from the CAD of unit 10 unless a measurement is named.

Frame​

PartSizeMaterial
Base plate306 x 306 x 10 mm, 2.4 kgEN AW-6082
Corner columns4 x item profile R30-90°, 133 mm (78 mm in the one-layer version)EN AW-6063
Intermediate plate306 x 306 x 6 mm, 1.35 kgEN AW-6082
Stage base plate310 x 310 x 12 mm, 2.1 kgEN AW-6082
X plate310 x 310 x 15 mm, 1.8 kgEN AW-6082
Y plate304 x 304 x 12 mm, 1.5 kgEN AW-6082
Sample plate (cover plate)310 x 310 x 8 mm, pocketed, 1.2 kgEN AW-5754
CladdingssheetEN AW-5005, anodised
Lamp arm, lamp shell, drawer guides, sensor flagsprintedPLA
Feet4 adhesive rubber feet (3M Bumpon SJ6125)

The unit stands on four small rubber feet and has no levelling screws. Printed PLA parts soften near 55 °C; the manual limits use to 38 °C.

A complete unit without optics modules has about 1,050 parts in CAD, of which about 580 are screws and standard parts. A circuit board counts as one part; its components are not counted.

XY stage​

Stage with the covers and the sample plate removed

Stage with the cover plate, claddings and lamp arm hidden. Axis directions: Coordinates and axes.

PropertyXY
Travel127 mm86 mm
Guides2 x HIWIN MGN9, rail 220 mm, two carriages per rail2 x HIWIN MGN9, rail 180 mm
Rail pairingone rail with zero clearance, one with a floating carriagesame
Carriage spacing along the travel48 mm48 mm
DriveDINGS 11E2110BS14 NEMA 11 motor, ball screw 6 mm diameter, 1 mm lead, grade C7same
Step5 µm per full step, 0.3125 µm per microstep (1/16)same
Motor1.0 A, 2.1 Ω per phasesame
Home sensorPanasonic PM-L25 fork light barrier with a printed flagsame
Feedbacknone as shipped; linear magnetic encoder (AS5311 type, 1.95 µm per count) in validationsame

The X axis sits on the stage base plate. The Y axis, both focus drives and their driver boards ride on the X plate and are fed through a drag cable.

The sample sits about 70 mm above the X rails. Rail straightness therefore shows up as a sideways error at the sample that an encoder at plate level does not see.

Measured behaviour is on the page Stage repeatability and drift.

Focus drive​

Focus lifts the sample plate; the objective is fixed. See Design decisions for the reason.

PropertyValue
Mechanismtwo LZ40 wedge stages (ratio 1:1, low spring force), 178 mm apart on the Y plate, at the front and back of the sample plate
Actuatorstwo DINGS L8K2105AD4 NEMA 8 captive actuators, one per wedge stage
Lead0.5 mm, plastic (POM) nut
Step2.5 µm per full step; 0.156 µm at 1/16 microstepping
Travel8 mm used (±4 mm); 12.7 mm stroke
Thrustabout 30 N per actuator
Lifted masssample plate 1.2 kg (CAD) plus holder and sample, about 1.5 to 1.8 kg
Home sensorfork light barrier
Feedbacknone as shipped; one rotary encoder planned
Check the Z step size of your unit

The example configuration lists 0.3125 µm per step for Z, the same as X and Y. With a 0.5 mm lead and a 1:1 wedge that value is right only if the Z drivers run at 1/8 microstepping. Verify on your unit by focusing on both faces of a cover glass of known thickness.

Do not lift the unit by the sample plate and do not load the plate by hand: the plate is held down by its weight and the stage springs only.

Known weaknesses of the focus drive​

  1. Two drives, one plate. If one side loses steps, the plate tilts. 100 full steps of difference tilt it by 1.4 mrad, which is 120 µm of focus across the Y travel.
  2. Both supports stand on one line (x = 0). A push on the left or right edge of the plate is resisted only by the roller guides of the two wedge stages.
  3. Downward motion relies on a weak spring and gravity. With the plastic nut this gives backlash and stick-slip on reversal.

How to measure the focus repeatability​

The camera's own focus measure is too coarse for this: the depth of field of a 20x/0.40 is several micrometres. Use a displacement measurement:

MethodResolutionWhat you need
Focus-lock module as a displacement sensor on a cover glass0.08 µm at 20x (0.75 µm per pixel, fitted to 0.1 pixel), range ±0.9 mmthe focus lock add-on
Digital length gauge on the sample plate0.1 µma gauge with a plunger force under 1 N, on a stand fixed to the stage base plate
Capacitive or confocal distance sensor0.01 to 0.05 µmlab sensor, no contact force
Tilted grid target: the line of best focus moves sideways by z / tan(tilt)about 0.3 µm at 10x with a 5° tilta grid slide on a 5° wedge

Procedure, following the logic of ISO 230-2:

  1. Warm up for 30 minutes with the motors energised.
  2. Pick five target heights across the travel.
  3. Approach each target seven times from below and seven times from above, with a 2 s pause before reading.
  4. Report for each target: the spread of the approaches from one side (unidirectional repeatability), and the difference between the two means (reversal error).
  5. Repeat with gauges at the front and back wedge stage at the same time. The difference between them is the tilt.
  6. Repeat with a loaded well plate and at three XY positions.

Planned improvements​

The sample keeps moving for focus. The changes address the three weaknesses above:

ChangeEffect
One encoder per focus actuator, and one home switch per side compared at every starta lost step on one side is detected before the plate tilts
Always approach the target from below, with an overshoot on downward movesremoves the reversal error from the images, with no hardware change
Stronger return springs or an anti-backlash nutremoves stick-slip on the way down
A third support, or one motor that drives both wedge stages through a shaftthe plate can no longer tilt by drive error
Piezo fine axis: one piezo stack per wedge stage, driven in tandem (two-channel driver, 0 to 138 V, 12 bit; board in design)fine focus without stepper noise or backlash; with a 100 µm stack about 25 nm per step
Light piezo insert that lifts only the sample holder instead of the whole platea few hundred grams of moving mass; fast z-stacks

A piezo stack without a position sensor shows 10 to 15 % hysteresis and creep. Close the loop with the focus lock or a strain gauge when absolute height matters.

Objective slider​

Objective slider

Stage base plate with the X and Y plates hidden. The objectives themselves are hidden; their supports sit on the slider.

PropertyValue
Positionstwo, 29.8 mm apart
Guideone HIWIN MGN7 carriage on a 57 mm rail
Drivesame NEMA 11 ball-screw motor as X and Y
Positioningby motor steps from a home sensor; no hard stop and no encoder
Objective mountthree-point mount held by two magnets; one support per objective sets the parfocal height
Largest objective diameterabout 28 mm
Crash detection (FR120, optional)thin-film force sensor under one support point of the mount

One full step moves the objective 5 µm, which is 42 camera pixels at 20x. Calibrate both positions on your unit; see Calibrate objective.

The force sensor is intended as a crash detector: it reports when the objective touches the sample plate or a holder. It does not replace a height limit per objective in software.

Optics drawer​

Optics drawer

The drawer with the stage, the intermediate plate and the columns hidden.

PropertyValue
Grid3 x 3 cubes of 50 mm, one or two layers of 55 mm height
Platesaluminium, 150 x 150 x 5 mm, top plate with a 35 mm port on the axis
Supportthree ball points (Ganter GN 605) on the base plate, at the front centre and at the left and right of the axis
Clampan inverted LZ40 stage with nine rubber buffers presses on the drawer from above, inside the triangle of the ball points; tightened from the front with a hex key
Side guides and end stopprinted, 8 mm high
Slot above the drawer33 mm high, for slide-in modules; entry from the side

Unplug the camera's USB cable before you pull the drawer. Push the drawer against its end stop before you clamp it; pressing it in while clamped can move it off the axis.

Lamp arm​

The lamp head hangs on a printed arm, 185 mm in front of its column, with its diffuser 38 mm above the sample plate. A fan sits on top of the LED heat sink. A hinged version of the arm exists.